Note
Go to the end to download the full example code.
Dual-energy CBCT#
Material decomposition from two scans at different tube voltages.
Source: MATLAB/Demos/d23_DualEnergy.m
This demo is only available in MATLAB, so it is shown without output.
Data loading (See d21_ScannerDataLoader.m)#
% We'll use VarianDataLoader as an example. Load two sets of scan data
% at different kV settings into memory. Geo should be the same for both
% scans, and so only needs to be loaded once. Beam hardening correction
% is not needed.
datafolder_low='~/your_data_path/varian/low_kV/2020-01-01_123456/';
datafolder_high='~/your_data_path/varian/high_kV/2020-01-01_123456/';
[proj_h, geo, angles_h] = VarianDataLoader(datafolder_high, 'bh', false);
[proj_l, ~, angles_l] = VarianDataLoader(datafolder, 'bh', false);
Material Decomposition#
% DE-CBCT decomposes the two scans into equivalent thicknesses of two basis
% materials. Generally, either Al and PMMA or Iodine and Water are used.
% The calibration in TIGRE has uses Al and PMM, but other materials are
% supported if the calibration files are updated with the data (contact
% developers or make a PR for expanding this)
%
% The decomposition procedure defaults to taking scans at 80 and 140 kV.
% Different energies can be specified using the 'kVl' and 'kVh' tags.
[Al_proj, PMMA_proj, angles] = DeDecompose(proj_l, angles_l, proj_h, angles_h, ... % The following are defaults
'scanner','VarianTrueBeam','material1','Al','material2', 'PMMA','kVl', 80, 'kVh', 140);
Image construction#
% Equivalent thickness projections can now be used to synthesize various
% DE-derived image types. The most common of these are virtual
% monoenergetic (VM) images.
energy = 60; % keV
VM_proj = MakeVMproj(Al_proj, PMMA_proj, energy,'material1','Al','material2','PMMA');
VMimg = FDK(VM_proj, geo, angles);
VMimg = OS_SART(VM_proj, geo, angles, 100);
Plot image#
plotImg(VMimg)